Product Overview: ThermoProbe TL1-W Intrinsically Safe Laboratory Thermometer
ThermoProbe TL1-W Intrinsically Safe Laboratory Thermometer
Obsolete
See the ThermoProbe TL3-W here.
The ThermoProbe TL1-W intrinsically safe thermometer is a modified field version of our successful TL1-A thermometer. It features a rugged frame for reinforcement and is designed for continuous operation, but can be easily powered off if desired.
The TL1-W thermometer is intended to be used anywhere a precision glass stem thermometer or other type thermometer may be used. The TL-1W incorporates glass thermometer simplicity and reliability with the ease of a digital display. Like precision mercury thermometers, the TL1-W has accuracy, repeatability and long-term stability. But unlike those thermometers there will never be problems with mercury separation or hazardous material cleanup. Different from most precision electronic thermometers, the ‘cordless’ TL1-W has no wires to get in the way or break. Battery replacement is easily done through a sealed cover. Encoded calibration is unaffected by temperature, vibration or battery removal. Low battery display, and failure mode indications protect the user from false data.
4 Point NIST Traceable Calibration and a one year warranty included.
KEY FEATURES
- Minimum reading
- Maximum reading
- Averaging
- Easy °F to °C switching
- Backlight
- Sturdy frame
- Display arrows indicate increasing or decreasing temperature
Specifications
Display: | LCD with 0.25 inch characters Trend Indicating Arrows Low Battery & Fault indication |
User Functions: | Auto-off after 20 minutes or continuous operation Lowest Reading, Highest Reading, Average Reading Switchable in Units of °F or °C Display tenths or hundredths resolution |
Construction: | Stainless steel temperature sensor; Standard Sensor length 8′ (20cm) Standard diameter 1/4′ (6.35mm) with reduced diameter tip of 3/16′ (4.76mm) Sturdy aluminum frame surrounding aluminum enclosure Electronics and display in 3.5′ x 1.2′ anodized Aluminum enclosure Molded Silicone seal. Highly durable display window, graphics and membrane switches. |
Weight: | 6.5 ounces, 180 grams |
Temperature: | Resolution 0.01 Degrees Calibrated Accuracy: ±0.1°F from 14 to 320°F ±0.06°C from -10 to 160°C NIST Traceable Report of Test |
Temperature Time Constant: | 8 seconds in circulated water bath (63.2% response to temperature change) |
Electronics: | Ultra Low Power Microcontroller FLASH Memory 20 bit A/D Converter Recommended Ambient Temperature Range for Best Accuracy 32° to 112°F, 0° to 44°C |
Sensor: | 100 Ohm, Class A 1/3 DIN RTD Immersion depth 4′, 11cm |
Battery: | 2 AAA Alkaline; Duracell MN2400 Battery Life several hundred hours depending on backlight usage Battery Operating Range -4 to 130°F, -20 to 54°C Note: Battery may not provide adequate power if ambient temperature is below -4°F, -20°C or above 130°F, 54°C |
Typical Applications
Distillations, temperature baths, melting point tests, flash point tests, reference, thermowells, provers, temperature calibration for flow meters. | |
Industries: | Laboratory Petroleum Pipeline Pharmaceutical Food |
Reviews