Product Overview: Fluke 719 Series Electric Pressure Calibrator
Calibrate pressure with one hand using the Fluke 719 Electric Pressure Calibrator
Forget the hassle of hand pumping. Now you can calibrate and test pressure devices quickly and easily with one hand, saving valuable time. With the innovative, built-in electric pump, the Fluke 719 Electric Pressure Calibrator provides pressure calibration at your fingertips!
Use the programmable pump limit settings to eliminate over-pressurization, and to set “pump to” pressure values. The unique cleanout port design reduces pump failures, and makes the 719 easy to clean without repair. The compact, lightweight form, combined with one-handed calibration, makes this pressure calibrator the ideal tool for busy process professionals.
The Fluke 719 delivers pressure calibration made easy, by a name you know and trust for all your process calibration needs.
Features:
-
Electric pump for one handed pressure calibration
-
Best in class pressure measurement uncertainty of 0.025 %
-
Measure and source mA with best in class 0.015 % accuracy
-
Precision pressure adjust vernier
-
Variable release rate bleed valve for easy pressure adjustment
-
Ideal performance for high accuracy transmitter calibration
-
100 PSI pressure range
-
Programmable pump limit:
-
Eliminates over-pressurization
-
Allows the user to program target pressures for pumping
-
-
Extended measurement range with any of twenty-nine (29) 700Pxx Pressure Modules
-
Source mA with simultaneous pressure measurement to test valves and I/Ps
-
Simulate mA signals to troubleshoot 4-20 mA loops
-
Test pressure switches with built in switch test function
-
Power transmitters during test using 24 V loop supply
-
Proven cleanout ports reduce pump failures
Pump Specifications:
-
Maximum pressure generation of 120 PSI (8 BAR) (100 PSI model)
-
Pump to:
-
100 psi in less than 45 seconds*
-
*(with 1 meter of 1/8 inch test hose)
-
-
5000 pump cycles between rebuilds
-
Total pump life 20,000 calibration cycle
General Specifications:
-
Range: -12 to 120 psi (850 mbar to 8 bar)
-
Resolution: 0.01 psi, 1 mbar
-
Accuracy: 0.025 % range for 6 months. 0.035 % of range forone (1) year
-
mA range: 0 to 24 mA
-
mA Resolution: 0.001 mA
-
mA Accuracy: .015 % Rdg + 2 counts
-
mA: Max load, 1000 Ω; 750 Ω ohm in HART mode
-
Loop Supply Range: 24 V dc
-
Loop Supply Accuracy: ± 10 %
-
Storage temperature: -30 °C to 60 °C
-
Operating temperature: -10 °C to 55 °C
-
Relative humidity: 95 % (-10 °C to 30 °C); 75 % (30 °C to 40 °C); 45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C)
-
Shock: 1 meter drop test
-
Safety: CSA C22.2 No. 1010.1:1992
-
EMC: EN50082-1:1992 and EN55022:1994 Class B
-
Size (HxWxL): 60 mm x 87 mm x 210 mm (2.19 in x 3.41 in x 8.28 in)
-
Weight: 912 g (2 lb)
-
Battery: Two 9 V alkaline batteries
-
Battery life: 12 hours with 12 mA into 500 Ω, 200 pump cycles to 100 psi, 500 pump cycles to 30 psi
-
Warranty: Three-years, one year for pump
See more Pressure Calibrators here.