Product Overview: Fluke 283 FC/PV Solar Digital Multimeter and Wireless Current Clamp
The Fluke 283 FC/PV Solar Digital Multimeter and Wireless Current Clamp kit is the ultimate testing solution for professionals working in high-voltage solar environments. Engineered to meet the demanding needs of large-scale solar PV arrays, wind power installations, electric railways, and data centers, this kit significantly enhances safety, productivity, and measurement accuracy.
The CAT III 1500 V and CAT IV 1000 V True-RMS ratings ensure maximum protection when working with high-voltage DC systems. The included Staubli MC4 test leads allow for secure DC voltage testing up to 1500 V, while the TL175-HV silicone leads offer versatility for various measurement scenarios. The a283 FC Wireless Current Clamp enables safe current measurements up to 60 A AC/DC, perfect for testing module strings without direct electrical contact. The multimeter’s backlit LCD and illuminated keypad ensure clear visibility in all lighting conditions, enhancing usability in diverse field environments. With features like polarity indicators and a user-defined limit gauge, technicians can quickly identify wiring errors and perform efficient go/no-go troubleshooting.
The multimeter’s self-check function provides peace of mind by confirming functionality before critical measurements. Leveraging Fluke Connect™ technology, users can log, save, and view measurement data remotely, streamlining documentation and analysis processes. Whether you’re commissioning new solar installations, performing maintenance, or troubleshooting complex high-voltage systems, the Fluke 283 FC/PV kit delivers the precision, safety, and reliability needed for professional solar and high-voltage DC testing.
Advanced Features
- User-Defined Limit Gauge: Alerts for measurement deviations to streamline troubleshooting.
- Multimeter Readiness Self-Check: Verifies calibration, battery life, and functionality.
- Wireless Current Clamp: Non-contact measurements up to 60 A, with wireless data transmission and slim jaw design for tight spaces.
- Simultaneous Measurement: Voltage and current with automatic VA power calculation.
- Logging and Saving: Real-time clock, internal memory, and Fluke Connect™ for data transfer and analysis.
- Fluke Connect™: Real-time data transmission, cloud storage, and collaborative reporting.
Function | Range | Resolution | Accuracy |
AC volts[1] | 6.000 V | 0.001 V | 1.0 % + 3 (45 Hz to 500 Hz) |
60.00 V | 0.01 V | ||
600.0 V | 0.1 V | 2.0 % + 3 (500 Hz to 1 kHz) |
|
1000 V | 1 V | ||
DC volts | 6.000 V | 0.001 V | 0.09 % + 3 |
60.00 V | 0.01 V | ||
600.0 V | 0.1 V | ||
1500 V | 1 V | 0.15 % + 2 | |
AC mV[1] | 600.0 mV | 0.1 mV | 1.0% + 3 (45 Hz ~ 500 Hz) 2.0% + 3 (500 Hz ~ 1000 Hz) |
DC mV | 600.0 mV | 0.1 mV | 0.09 % + 2 |
Continuity | 600 Ω | 1 Ω | |
Resistance | 600.0 Ω | 0.1 Ω | 0.5 % + 4 |
6.000 kΩ | 0.001 kΩ | ||
60.00 kΩ | 0.01 kΩ | ||
600.0 kΩ | 0.1 kΩ | ||
6.000 MΩ | 0.001 MΩ | ||
50.00 MΩ | 0.01 MΩ | 1.5 % + 4 | |
Capacitance | 1000 nF | 1 nF | 1.2 % + 2 |
10.00 μF | 0.01 μF | ||
100.0 μF | 0.1 μF | ||
9999 μF | 1 μF | 10 % typical | |
Frequency | 99.99 Hz | 0.01 Hz | 0.1 % + 2 |
999.9 Hz | 0.1 Hz | ||
9.999 kHz | 0.001 kHz | ||
99.99 kHz | 0.01 kHz | ||
AC VA (45~500Hz)[1] | 360.0 VA | 0.1 VA | 2 % + 1.0 VA |
3.600 kVA | 0.001 kVA | 2 % + 0.01 kVA | |
36.00 kVA | 0.01 kVA | 2 % + 0.1 kVA | |
60.00 kVA | 0.01 kVA | 2 % + 0.15 kVA | |
DC VA[1] | 360.0 VA | 0.1 VA | 2 % + 1.0 VA |
3.600 kVA | 0.001 kVA | 2 % + 0.01 kVA | |
36.00 kVA | 0.01 kVA | 2 % + 0.1 kVA | |
90.00 kVA | 0.01 kVA | 2 % + 0.25 kVA | |
Min-max avg | For DC functions, accuracy is the specified accuracy of the measurement function ± 12 counts for changes longer than 350 ms in duration. | ||
For AC functions, accuracy is the specified accuracy of the measurement function ± 40 counts for changes longer than 900 ms in duration. |
[1] <1% range, unspecified
Environmental specifications | |
Operating temperature | -10 °C to 60 °C |
Storage temperature | -30 °C to +70 °C |
Humidity (without condensation) | 0 % to 90 %, 10 °C to 30 °C |
0 % to 75 %, 30 °C to 40 °C | |
0 % to 45 %, 40 °C to 60 °C |
Warranty and Protection | |
Electromagnetic compatibility | IEC 61326-1, IEC 61326-2-2, Portable, Group 1, Class A |
Overvoltage category | CAT III 1500 V, CAT IV 1000 V |
Agency approvals | ETL (AMER and EMEA), CSA (APAC), CE, UK CA |
Drop test | 2 meter (6.5 ft) drop test |
Ingress protection | IP52 |
Warranty | Limited lifetime |
Mechanical and general specifications | |
Counts | 6000 |
Size | 22.5 cm x 10.5 cm x 5.7 cm (8.8 in x 4.1 in x 2.2 in) |
Weight | 0.7 kg (1.5 lb) |
Battery life | > 150 hours typical, without backlight (Alkaline, 3 AA) |
> 100 hours typical when connected to wireless current clamp (Alkaline, 3 AA) |
a283 FC Wireless Current Clamp
Function | Range | Resolution | Accuracy |
AC current (True-RMS)[2] | 60 A | 0.01 A | 1.5% + 0.15 A, 45~500Hz |
DC current[2] | 60 A | 0.01 A | 1.5% + 0.15 A |
General specifications – Fluke a283 FC Wireless Current Clamp | |
Operating temperature | -10 °C to 60 °C |
Storage temperature | -30 °C to +70 °C |
Humidity (without condensation) | 0 % to 90 %, 5 °C to 30 °C |
0 % to 75 %, 30 °C to 40 °C | |
0 % to 45 %, 40 °C to 60 °C |
Warranty and Protection | |
Electromagnetic compatibility | IEC 61326-1, IEC 61326-2-2, Portable, Group 1, Class A |
Overvoltage category | CAT III 1500 V, CAT IV 1000 V |
Agency approvals | ETL (AMER and EMEA), CSA (APAC), CE, UK CA |
Drop test | 2 meter (6.5 ft) drop test |
Ingress protection | IP52 |
Warranty | 1-year |
Mechanical and general specifications | |
Size | 226.0 mm x 91.0 mm x 42.0 mm (8.9 in x 3.6 in x 1.7 in) |
Jaw opening | 34 mm (1.34 in) |
Weight | 0.375 kg (13.2 oz) |
Battery life | > 80 hours typical (Alkaline, 2 AAA) |
Features | |
LED indicator | Indicates wireless connection and data transmission status |
Battery indicator | Indicates battery status |
Power Key | Power On/Off |
Zero Key | A DC zeroing function |
[2] <1% range, unspecified