Product Overview: Additel 223A Documenting Process Calibrator
ADT 223A
- Sourcing, simulating and measuring pressure, temperature and electrical signals
- Smartphone-like menu and interface make the operation simpler and easier
- The internal cold junction compensation sensor can be re-calibrated at ice point by users
- Ultra-compact, 3.9” x 7.6” x 2.0”, and 1.6 lb (0.7 kg)
A highly integrated multifunction calibrator featuring several patented technologies, the 222A is an ultra-compact, rugged, and easy to use hand-held device for sourcing, simulating and measuring pressure, temperature, and electrical signals. Its smartphone-like menu and interface make the operation simpler and easier. The 222A is ideal for calibrating, maintaining, and troubleshooting process instrumentations. Automation and documentation capabilities make the 222A a turnkey solution.
Features
• Sourcing, simulating and measuring temperature and electrical signals
Sources and measures mV, mA, ohms, RTDs, thermocouples, frequency, and pulses
Simulates and measures 13 thermocouples and 11 RTDs to calibrate transmitters
Measures and sources pressure using 160 series Intelligent Digital Pressure Modules from -15 psi to 10,000 psi (-1 bar to 700 bar)
Measures and sources pressure using 160 series Intelligent Digital Pressure Modules from -15 psi to 10,000 psi (-1 bar to 700 bar)
24V loop power supply
Simultaneous dual reading capability
Automatic switch test
Supports square root transmitter
Pulse frequency output for the calibration of flow totalizer
• Easy to use
Smartphone-like menu and interface make the operation simpler and easier
Ultra-compact, size 3.9” x 7.6” x 2.0” (100mm x 192mm x 52mm), and weight 1.6 lb (0.7 kg)
One hand operation
• Calibrated cold junction compensation (Patented)
Cold junction equivalent block in the calibrator
A calibrated PRT element with flexible leads is installed in the equivalent block for thermocouple cold junction compensation
This PRT element can be pulled out from the calibrator and re-calibrated and corrected at ice point by users
• Built-in temperature readout
CVD coefficients of a calibrated PRT can be input into the calibrator for accurate temperature measurement.
• Multi lingual interface
English, German, French, Italian, Spanish, Portuguese, Simplified Chinese
(Traditional Chinese, Japanese and Russian are available per request)
(Traditional Chinese, Japanese and Russian are available per request)
• Documenting and automated procedure capability
Manage the information of the device under test.
Set up automated calibration procedures, and ADT 221A performs the test, calculates the errors, displays and/or stores the results in the memory, and highlights the out-of-tolerance points.
As-found and As-left functions allow recording and documenting results for quality control.
Download tasks and upload the results.
Snapshots allow you to capture and save work.
• Hart Communication capability
Support HART® instrumentation
• Build-in unit conversion tool
Build-in converters for pressure units, temperature units, temperature vs. resistance (RTDs), and temperature vs millivolt (thermocouples)
• Display
3.5 inch TFT color screen
• Rugged
Rugged design for harsh environments.
Passed a 1-meter drop test.
Three year warranty for the ADT 221A, and one year for the battery pack
• Misuse protection
Up to 30V voltage on any two sockets and up to 1A current on current sockets will not damage the calibrator. The calibrator will return to normal condition as soon as the voltage or current is removed.
• Rechargeable battery
Rechargeable Li-ion battery for 15 hours uninterrupted use.
Battery life will be reduced when 24V is applied.
The rechargeable battery is replaceable.
• NIST Traceable Calibration with data
• Warranty
1 year.
Applications
The 222A multifunction calibrator is a process tool for measuring, sourcing and simulating mA, mV, V, RTDs, thermocouples, ohms, frequency, pulses and pressure, captures switch values and provides 24 V loop power.
• Electrical
Resistance measurement / simulation
Voltage measurement / generation
Current measurement /generation
Frequency measurement / generation
Pulse counting/generation
24 VDC loop supply
Switch sensing
• Temperature
RTD measurement /simulation
TC measurement /simulation
• Pressure
Gauge pressure measurement
Absolute pressure measurement
Absolute pressure measurement
Specifications
• Electrical Specifications
Measurement Accuracy
|
|||||
Range
|
Resolution
|
Accuracy
|
|||
Voltage DC
|
±75.0000 mV
±30.0000 V
|
1µV
0.1mV
|
0.01%RD+0.005%FS
0.01%RD+0.005%FS
|
||
Current DC
|
±30.0000 V
|
0.1µA
|
0.01%RD+0.005%FS
|
||
Resistance
|
Two-wire
|
0 to 400.000 Ω
|
1mΩ
|
0.02%RD+0.005%FS
|
|
Three-wire
|
0 to 400.000 Ω
|
1mΩ
|
0.02%RD+0.005%FS
|
||
Four-wire
|
0 to 400.000 Ω
|
1mΩ
|
0.01%RD+0.005%FS
|
||
Two-wire
|
0 to 400.000 Ω
|
10mΩ
|
0.02%RD+0.005%FS
|
||
Three-wire
|
0 to 400.000 Ω
|
10mΩ
|
0.02%RD+0.005%FS
|
||
Four-wire
|
0 to 400.000 Ω
|
10mΩ
|
0.01%RD+0.005%FS
|
||
Frequency
|
1 to 50000.0 Hz
|
0.1Hz
|
0.005%RD+0.002%FS
|
||
Pulse
|
0 to 999999
|
1
|
N/A
|
||
Limit Switch
|
If the switch has detective voltage, its range is +3V to +24V
|
||||
Source Accuracy
|
|||||
Voltage DC
|
-10.000 to 75.000mV 0 to 12.0000 V
|
1μV
0.1mV
|
0.02%RD+0.005%FS
0.02%RD+0.005%FS
|
||
Current DC
|
0 to 22.000 mA
|
1µA
|
0.02%RD+0.005%FS
|
||
Resistance
|
1 to 400.00 Ω
1 to 4000.0 Ω
|
10mΩ
100mΩ
|
0.02%RD+0.005%FS
0.03%RD+0.01%FS
|
||
Frequency
|
0 to 50000.0 Hz
|
0.1Hz
|
0.005%RD+0.002%FS
|
||
Pulse
|
0 to 999999
|
1
|
N/A
|
||
DC24V
|
N/A
|
N/A
|
0.5V
|
General Specifications
Environmental Specifications
|
|
Operating Temperature
|
-10°C to 50°C
|
Storage Temperature
|
-20°C to 60°C
|
Humidity
|
<90%, non-condensing
|
Safety Specifications
|
|
European Compliance
|
CE Mark
|
Mechanical Specifications
|
|
Display
|
3.5 inch TFT color screen
|
Electrical Connection
|
φ4mm sockets and flat mini-jack thermocouple socket
|
RS232 Interface
|
standard RS232-DB9 socket
|
Size
|
3.9” x 7.6” x 2.0” (100mm x 192mm x 52mm)
|
Weight
|
1.6 lb (0.7 kg)
|
Power Supply
|
Polymer Li-ion rechargeable battery, or 10V DC adaptor
|
Battery
|
Rechargeable Li-ion battery (included)
|
Battery Life
|
15 hours uninterrupted use
Battery life will be reduced when 24V is applied
|
Battery Charge
|
110V/220V external power adapter (included)
|
Pressure Specification
|
|
ADT 160 series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.025% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detail specification refer to pressure modules datasheet.
|
• Temperature Specifications
Thermocouple Measurement and Source accuracy
|
||||||||||
Measure
and Simulate
|
Standard
|
Temperature Range (°C)
|
Accuracy (°C)
|
|||||||
Measure
|
Source
|
|||||||||
S
|
IEC 584
|
-50 to 1768
|
-50 to400
|
1.0
|
1.1
|
|||||
400 to 1000
|
0.6
|
0.6
|
||||||||
1000 to1768
|
0.7
|
0.8
|
||||||||
R
|
IEC 584
|
-50 to 1768
|
-50 to 200
|
1.4
|
1.4
|
|||||
200 to 500
|
0.6
|
0.6
|
||||||||
500 to 1768
|
0.6
|
0.7
|
||||||||
B
|
IEC 584
|
0 to 1820
|
50 to 450
|
3.8
|
3.8
|
|||||
450 to 800
|
0.9
|
0.9
|
||||||||
800 to 1820
|
0.6
|
0.7
|
||||||||
K
|
IEC 584
|
-270 to 1372
|
-250 to -200
|
1.0
|
1.1
|
|||||
-200 to -100
|
0.4
|
0.5
|
||||||||
-100 to 600
|
0.3
|
0.3
|
||||||||
600 to 1372
|
0.4
|
0.5
|
||||||||
N
|
IEC 584
|
-270 to 1300
|
-250 to -200
|
1.5
|
1.6
|
|||||
-200 to -100
|
0.5
|
0.6
|
||||||||
-100 to 1300
|
0.4
|
0.5
|
||||||||
E
|
IEC 584
|
-270 to 1000
|
-250 to -200
|
0.6
|
0.7
|
|||||
-200 to -100
|
0.3
|
0.3
|
||||||||
-100 to 0
|
0.2
|
0.2
|
||||||||
0 to 700
|
0.2
|
0.3
|
||||||||
700 to 1000
|
0.2
|
0.4
|
||||||||
J
|
IEC 584
|
-270 to 1200
|
-210 to -100
|
0.3
|
0.3
|
|||||
-100 to 1200
|
0.3
|
0.4
|
||||||||
T
|
IEC 584
|
-270 to 400
|
-250 to -200
|
0.8
|
0.9
|
|||||
-200 to 0
|
0.4
|
0.4
|
||||||||
0 to 400
|
0.2
|
0.2
|
||||||||
C
|
ASTM E988
|
0 to 2315
|
0 to 1000
|
0.5
|
0.5
|
|||||
1000 to 1800
|
0.7
|
0.9
|
||||||||
1800 to 2315
|
1.0
|
1.4
|
||||||||
D
|
ASTM E988
|
0 to 2320
|
0 to 100
|
0.5
|
0.5
|
|||||
100 to 1100
|
0.4
|
0.5
|
||||||||
1100 to 2000
|
0.6
|
0.9
|
||||||||
2000 to 2320
|
0.9
|
1.3
|
||||||||
G
|
ASTM E1751
|
0 to 2315
|
0 to 200
|
2.4
|
2.4
|
|||||
200 to 400
|
0.5
|
0.5
|
||||||||
400 to 1400
|
0.4
|
0.5
|
||||||||
1400 to 2315
|
0.7
|
1.0
|
||||||||
L
|
DIN43710
|
-200 to 900
|
-200 to -100
|
0.2
|
0.3
|
|||||
-100 to 400
|
0.2
|
0.2
|
||||||||
400 to 900
|
0.2
|
0.3
|
||||||||
U
|
DIN43710
|
-200 to 600
|
-200 to 0
|
0.4
|
0.4
|
|||||
0 to 600
|
0.2
|
0.3
|
||||||||
RTD Measurement and Source accuracy
|
||||||||||
Measure
and Simulate
|
Standard
|
Temperature Range (°C)
|
Accuracy (°C)
|
|||||||
Measure (2W/3W)
|
Measure (4W)
|
Source
|
||||||||
Pt10(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.65
|
0.60
|
0.65
|
||||
200 to 600
|
0.82
|
0.72
|
0.82
|
|||||||
600 to 850
|
0.96
|
0.82
|
0.96
|
|||||||
PT100
(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.15
|
0.1
|
0.15
|
||||
200 to 600
|
0.26
|
0.16
|
0.26
|
|||||||
600 to 850
|
0.34
|
0.20
|
0.34
|
|||||||
Pt100
(3916)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.15
|
0.1
|
0.15
|
||||
200 to 600
|
0.26
|
0.16
|
0.26
|
|||||||
600 to 850
|
0.33
|
0.20
|
0.33
|
|||||||
Pt200(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.37
|
0.32
|
0.69
|
||||
200 to 600
|
0.51
|
0.41
|
0.92
|
|||||||
600 to 850
|
0.61
|
0.48
|
1.08
|
|||||||
Pt500(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.20
|
0.16
|
0.36
|
||||
200 to 600
|
0.32
|
0.22
|
0.54
|
|||||||
600 to 850
|
0.40
|
0.27
|
0.67
|
|||||||
Pt1000
(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
Reviews