FLUKE 9100S

The 9100S Dry-Well sets the standard for portable dry-wells, outperforming others in its class despite its compact size (2¼ inches high and 5 inches wide) and lightweight design. It’s user-friendly, allowing anyone to master it in under 15 minutes. With a temperature range up to 375°C (707°F), it’s ideal for field testing RTDs, thermocouples, and small bimetal thermometers.

View Quote List
QTY:
 

Product Overview: Fluke Calibration 9100S Handheld Dry-Well

9100S Dry-Well

data sheet

The 9100S Dry-Well sets the standard for portable dry-wells, outperforming others in its class despite its compact size (2¼ inches high and 5 inches wide) and lightweight design. It’s user-friendly, allowing anyone to master it in under 15 minutes. With a temperature range up to 375°C (707°F), it’s ideal for field testing RTDs, thermocouples, and small bimetal thermometers.

Simply plug it in, turn it on, set the desired temperature, and insert your probe into the correctly sized well. Compare the device reading to the display temperature or an external reference to determine the error. Featuring a Fluke Calibration temperature controller, the 9100S offers 0.1°C resolution, ±0.5°C accuracy, ±0.1°C stability, and a quick stabilization time of five minutes.

Fluke Calibration 9100S Fixed-Block Options

Specifications 9100S 9102S
Range 35°C to 375°C (95°F to 707°F) –10°C to 122°C (14°F to 252°F) at 23°C ambient
Accuracy ±0.25°C at 50°C; ±0.25°C at 100°C; ±0.5°C at 375°C ±0.25°C
Stability ±0.07°C at 50°C; ±0.1°C at 100°C; ±0.3°C at 375°C ±0.05°C
Well-to-well uniformity ±0.2°C with sensors of similar size at equal depths within wells
Heating times 35°C to 375°C: 9.5 minutes Ambient to 100°C: 10 minutes
Stabilization 5 minutes 7 minutes
Cooling times 375°C to 100°C: 14 minutes Ambient to 0°C: 10 minutes
Well depth 102 mm (4 in); 1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep 102 mm (4 in)
Removable inserts N/A Available in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included]
Power 115 V AC (±10%), 55–65 Hz, 1.5 A or 230 V AC (±10%), 0.8 A, 45–55 Hz, 175 W 94-234 V AC (±10%), 50/60 Hz, 60 W; or 12 V DC
Size (H x W x D) 57 x 125 x 150 mm (2.25 x 4.9 x 6.1 in) 99 x 140 x 175 mm (3.9 x 5.5 x 6.9 in)
Weight 1 kg (2 lb 3 oz) 1.8 kg (4 lb)
Computer interface RS-232 included
NIST-traceable calibration Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C and 375°C Data at –10°C, 0°C, 25 °C, 50°C, 75 °C,100°C and 122°C